Add to Cart
OPC-P2 Portable Oil Particle Counter
OPC-P2 Portable Oil Particle Counter adopts laser extinction technology, is designed to measure the size and quantity of particles in oil liquid. It is suitable for testing the contamination level of solid particles in hydraulic oil, lubricating oil, fireproof oil, insulating oil, turbine oil, organic liquid and polymer solution, etc. So our instrument can be used in aviation, aerospace, electric power, petroleum, chemical industry, transportation, port, metallurgy, machinery, automobile manufacturing and many other fields.4.1 Power Choice: External power or battery.
* Built-in multiple calibration curves, compatible with all common domestic and international standards for calibration.
* Built-in NAS1638, GJB420A, GJB420B, ISO4406, SAE4059DPC, SAE4059CPC, GB/T14039, GOST17216, QC/T29104, JB/T9737, DL/T432, SAE749D, HH005-2018 and other common standards.
* One test can give the data results under all the built-in standards, support customized standard test, and can set the required standards according to customer needs;
* 1000 particle size channels can be set to facilitate particle size analysis
All the cases require the portable rapid detection of gas concentrations such as petroleum, chemical, pharmaceutical, environmental protection, combustion gas distribution, warehousing, smoke gas analysis, air governance, and so on.
Light source | Semiconductor laser |
Particle size range | 0.8-600μm (depending on different sensors) |
Sensitivity | 0.1μm |
Customized detection channels | 8~64 channels optional, arbitrary setting of particle size size |
Sampling volume | 0.2-1000ml, interval 0.1mL |
Sampling accuracy | better than ± 1% |
Sampling speed | 5-80mL/min |
Cleaning speed | 5-80mL/min |
Cleaning volume | can be set between 0~90mL |
Counting accuracy | <±5% |
Resolution | ≤ 10% |
Repeatability | RSD<2% |
Limit overlap error | 12000-40000 grains/mL |
Offline detection viscosity | ≤100cSt (viscosity is too large optional air pressure chamber) |
Pressure range | low pressure 0-0.6Mpa, high pressure up to 42Mpa (optional pressure reducing valve) |
On-line detection interval time | arbitrary settings |
Testing sample temperature | 0 ℃ ~ 80 ℃ |
Working temperature | -20℃~60 |
Storage temperature | -30℃~80 |
Screen size | 7 inch |
Power supply | AC100-240V, 50/60Hz |
Battery capacity | 5200mAh |
Battery running time | 6-8 hours |
Overall Dimension | 310×305×135mm |
Instrument weight | 7.5kg |